share:
delim
delim
Glycol Ethers of Regulatory Importance on Rxi-1301Sil MS

Glycol Ethers of Regulatory Importance on Rxi-1301Sil MS

PeakstR (min)Conc.
(µg/mL)
Common Name
1.EGME2.483100Ethylene glycol methyl ether
2.EGEE3.218100Ethylene glycol ethyl ether
3.Perfluoro TEGME (IS)3.894100Perfluoro triethylene glycol methyl ether
4.PnPGE4.412100Propylene glycol propyl ether
5.PGMA4.622100Propylene glycol methyl ether acetate
6.EGBE4.998100Ethylene glycol butyl ether
7.PGBE5.218100Propylene glycol butyl ether
8.DEGME5.320100Diethylene glycol methyl ether
9.DPGME I5.685100Dipropylene glycol methyl ether
10.DPGME II5.715Dipropylene glycol methyl ether
11.DEGEE5.763100Diethylene glycol ethyl ether
12.DPGME III5.807Dipropylene glycol methyl ether
13.1,2-DCB-D4 (IS)5.8761001,2-Dichlorobenzene-D4
14.EGHE6.383100Ethylene glycol hexyl ether
15.DEGBE6.907100Diethylene glycol butyl ether
16.EGPhE7.239100Ethylene glycol phenyl ether
17.TPGME isomers7.451100Tripropylene glycol methyl ether
18.DEGHE7.952100Diethylene glycol hexyl ether
Standard was prepared from commercially available neat compounds, 95-99% purity.
image
GC_PC1280
ColumnRxi-1301Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 16094)
Sample
Diluent:Methanol
Conc.:100 ppm
Injection
Inj. Vol.:1 µL split (split ratio 30:1)
Inj. Temp.:260 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 300 °C at 27 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:1.3 mL/min
Linear Velocity:41.05 cm/sec @ 40 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1220-22029
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:2 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Glycol Ethers

,

Solvents

,

Rxi®-1301Sil MS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Email

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us