Biodiesel (B100) on MXT-Biodiesel TG by ASTM D6584

GC_PC1293
Column | MXT-Biodiesel TG with 2 m Integra-Gap (integrated retention gap), 16 m total length, 0.53 mm ID, 0.16 µm (cat.# 70289) |
Sample | Biodiesel (B100), derivatized |
Diluent: | Heptane |
Injection | |
Inj. Vol.: | 1 µL cold on-column |
Temp. Program: | Oven track |
Oven | |
Oven Temp.: | 50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C at 30 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 3 mL/min |
Detector | FID @ 380 °C |
Instrument | Agilent 7890A GC |