Alkylate Standard on Rtx-DHA 50 by Method D5134-13
*Extrapolated from n-C6 and n-C7 or from n-C8 and n-C9

GC_PC1319
Column | Rtx-DHA-50, 50 m, 0.20 mm ID, 0.50 µm (cat.# 10147) |
Sample | ASTM D5134 qualitative reference alkylate standard (DCG Partnership I, LTD) |
Injection | |
Inj. Vol.: | 0.2 µL split (split ratio 200:1) |
Liner: | Premium 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305.5) |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 35 °C (hold 30 min) to 200 °C at 2 °C/min |
Carrier Gas | He, constant pressure |
Linear Velocity: | 23 cm/sec |
Detector | FID @ 250 °C |
Make-up Gas Flow Rate: | 30 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 370 mL/min |
Instrument | Agilent 7890B GC |
Notes | Understanding the hydrocarbon components in petroleum samples is useful for sample characterization, determining product quality, planning and modeling of refinery processes, and also for regulatory purposes. ASTM D5134-13 provides the composition of paraffin, naphthenes, and monoaromatic compounds up to n-C9 in petroleum naphthas, reformates, and alkalytes. This method is run using a long, narrow-bore column, which has a limited sample loading capacity. Therefore, high split is used and a small sample volume is injected. Restek’s Rtx-DHA-50 column is a good match for this analysis; retention index (RI) data are in agreement with the RI data provided in the method. |