share:
delim
delim
Aromatics on Rtx-DHA-100 (ASTM D6729-14)

Aromatics on Rtx-DHA-100 (ASTM D6729-14)

PeakstR (min)Conc.
(wt.%)
1.Benzene44.1648.255
2.Toluene67.51912.342
3.Ethylbenzene83.5348.248
4.m-Xylene84.7353.108
5.p-Xylene84.8811.041
6.o-Xylene87.7353.125
7.Isopropylbenzene91.9432.085
8.n-Propylbenzene95.4193.091
9.1-Methyl-3-ethylbenzene96.2742.795
10.1-Methyl-4-ethylbenzene96.5042.070
11.1,3,5-Trimethylbenzene97.1270.422
12.1-Methyl-2-ethylbenzene98.2071.571
13.1,2,4-Trimethylbenzene99.8212.066
14.tert-Butylbenzene99.8212.058
15.Isobutylbenzene101.5033.099
16.sec-Butylbenzene101.7873.090
17.1-Methyl-3-isopropylbenzene102.7951.066
18.1-Methyl-4-isopropylbenzene103.1584.126
19.1-Methyl-2-isopropylbenzene104.4672.806
20.1-Methyl-3-n-propylbenzene105.7502.084
21.1-Methyl-4-n-propylbenzene106.1332.073
22.n-Butylbenzene106.2282.077
23.1,3-Dimethyl-5-ethylbenzene106.3560.407
24.1-Methyl-2-n-propylbenzene107.2092.148
25.1,4-Dimethyl-2-ethylbenzene108.0812.076
26.1,2-Dimethyl-4-ethylbenzene108.7520.915
27.1,2-Dimethyl-3-ethylbenzene110.4592.152
28.1,3-Dimethyl-2-ethylbenzene110.4590.799
29.1,2,4,5-Tetramethylbenzene111.3980.860
30.(2-Methylbutyl)benzene111.5600.118
31.1-tert-Butyl-2-methylbenzene112.7052.077
32.Pentylbenzene114.3282.058
33.1-tert-Butyl-4-ethylbenzene116.0122.076
34.1,3,5-Triethylbenzene118.3342.087
35.1,2,4-Triethylbenzene119.4881.014
36.Hexylbenzene120.7165.155
37.1-tert-Butyl-3,4,5-trimethylbenzene123.9993.355
image
GC_PC1323
ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
SampleDHA aromatics standard (cat.# 30729)
Injection
Inj. Vol.:0.2 µL split (split ratio 200:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:0 °C (hold 15 min) to 50 °C at 1 °C/min to 130 °C at 2 °C/min to 270 °C at 4 °C/min
Carrier GasHe, constant pressure
Linear Velocity:24 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
InstrumentAgilent 7890A GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

DHA

,

PiONA

,

PONA

,

gasoline

,

D6729

,

D6730

,

D6733

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us