Naphthenes on Rtx-DHA-100 (ASTM D6729-14)
This chromatogram may contain analytes that are synonyms of your search terms.

Naphthenes on Rtx-DHA-100 (ASTM D6729-14)

PeakstR (min)Conc.
(wt.%)
1.Cyclopentane26.3699.455
2.Methylcyclopentane39.1975.921
3.Cyclohexane45.9874.773
4.cis-1,3-Dimethylcyclopentane51.7760.300
5.trans-1,3-Dimethylcyclopentane52.3605.799
6.trans-1,2-Dimethylcyclopentane52.9693.653
7.Methylcyclohexane59.8523.677
8.Ethylcyclopentane62.6717.837
9.ctc-1,2,3-Trimethylcyclopentane66.2863.738
10.trans-1,4-Dimethylcyclohexane72.1354.705
11.1-Ethyl-1-methylcyclopentane74.7041.459
12.trans-1,2-Dimethylcyclohexane75.1281.642
13.ccc-1,2,3-Trimethylcyclopentane76.0920.795
14.Isopropylcyclopentane77.4971.305
15.cis-1,2-Dimethylcyclohexane80.07010.276
16.1,1,4-Trimethylcyclohexane80.8786.304
17.n-Propylcyclopentane82.0524.373
18.ctt-1,2,4-Trimethylcyclohexane82.0521.839
19.ccc-1,3,5-Trimethylcyclohexane83.8852.343
20.ctc-1,2,3-Trimethylcyclohexane86.9923.676
21.ctc-1,2,4-Trimethylcyclohexane87.2111.062
22.1,1,2-Trimethylcyclohexane87.8351.842
23.Isobutylcyclopentane89.5331.334
24.Isobutylcyclohexane100.3363.666
25.tert-1-Methyl-2-N-propylcyclohexane101.4534.637
26.tert-1-Methyl-2(methylpentyl)cyclopentane114.5433.592
image
GC_PC1325
ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
SampleDHA naphthenes standard (cat.# 30728)
Injection
Inj. Vol.:0.2 µL split (split ratio 200:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:0 °C (hold 15 min) to 50 °C at 1 °C/min to 130 °C at 2 °C/min to 270 °C at 4 °C/min
Carrier GasHe, constant pressure
Linear Velocity:24 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
InstrumentAgilent 7890A GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

DHA

,

PiONA

,

PONA

,

gasoline

,

D6729

,

D6730

,

D6733

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us