share:
delim
delim
Simulated Distillation of Gasoline with Oxygenates on MXT-1 (ASTM D7096-16)

Simulated Distillation of Gasoline with Oxygenates on MXT-1 (ASTM D7096-16)

PeakstR (min)Conc.
(vol.%)
1.Methanol1.0894.1
2.Ethanol1.3024.1
3.tert-Butanol1.6704.1
4.MTBE2.0394.1
5.n-Hexane2.3296.1
6.ETBE2.4674.1
7.2,4-Dimethylpentane2.5856.1
8.TAME2.9534.1
9.n-Heptane3.19210.2
10.Toluene3.72212.2
11.n-Octane4.0465.1
12.p-Xylene4.5858.2
13.n-Propylbenzene5.2335.1
14.n-Decane5.5694.1
15.n-Butylbenzene5.9644.1
16.n-Dodecane6.8764.1
17.n-Tridecane7.4642.0
18.n-Tetradecane8.0152.0
19.n-Pentadecane8.5332.0
20.n-Hexadecane9.0234.1
image
GC_PC1327
ColumnMXT-1, 30 m, 0.53 mm ID, 5.00 µm (cat.# 70179)
SampleD7096 custom standard with oxygenates
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 245 °C at 25 °C/min (hold 4 min)
Carrier GasHe, flow program
Flow Program:5 mL/min (hold 0.5 min) to 20 mL/min at 60 mL/min/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:30 mL/min
Make-up Gas Type:He
Hydrogen flow:40 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Determination of the Boiling Range Distribution

,

gasoline

,

oxygenates

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us