share:
delim
delim
PiONA Calibration Standard on Rtx-DHA-100 by ASTM D6729-14
This chromatogram may contain analytes that are synonyms of your search terms.

PiONA Calibration Standard on Rtx-DHA-100 by ASTM D6729-14

PeakstR (min)Conc.Wt. %Group
1.3-Methyl-1-butene12.8050.291O5
2.iso-Pentane14.3140.304I5
3.1-Pentene15.891.669O5
4.2-Methyl-1-butene16.7120.42O5
5.n-Pentane17.2051.728P5
6.2-Methyl-1,3-butadiene17.7140.699O5
7.trans-2-Pentene18.2950.542O5
8.cis-2-Pentene19.2940.21O5
9.4-Methyl-1-pentene25.7160.646O6
10.Cyclopentane26.3581.891N5
11.2,3-Dimethylbutane27.1060.447I6
12.2-Methylpentane28.1591.211I6
13.3-Methylpentane30.7511.998I6
14.1-Hexene32.1392.137O6
15.n-Hexane34.441.729P6
16.trans-2-Hexene35.5660.31O6
17.2-Methyl-2-pentene36.0470.654O6
18.cis-2-Hexene37.360.555O6
19.Methylcyclopentane39.1921.184N6
20.2,2-Dimethylpentane39.2780.463I7
21.2,4-Dimethylpentane40.7831.127I7
22.2,2,3-Trimethylbutane41.3120.125I7
23.Benzene44.1411.899A6
24.3,3-Dimethylpentane45.7070.438I7
25.Cyclohexane45.980.954N6
26.2-Methylhexane49.1430.891I7
27.2,3-Dimethylpentane49.1430.887I7
28.3-Methylhexane50.9630.451I7
29.cis-1,3-Dimethylcyclopentane51.7860.06N7
30.trans-1,3-Dimethylcyclopentane52.3461.16N7
31.trans-1,2-Dimethylcyclopentane52.9650.73N7
32.3-Ethylpentane53.1410.168I7
33.1-Heptene53.9411.29O7
34.trans-3-Heptene56.1810.712O7
35.n-Heptane56.5041.723P7
36.cis-3-Heptene56.7220.194O7
37.trans-2-Heptene57.6460.101O7
38.cis-2-Heptene59.2990.649O7
39.Methylcyclohexane59.850.735N7
40.2,2-Dimethylhexane61.270.149I8
41.Ethylcyclopentane62.6421.567N7
42.2,2,3-Trimethylpentane63.3050.698I8
43.2,5-Dimethylhexane63.581.112I8
44.2,4-Dimethylhexane63.9120.276I8
45.ctc-1,2,3-Trimethylcyclopentane66.2820.747N8
46.Toluene67.462.839A7
47.2-Methylheptane70.6550.899I8
48.4-Methylheptane70.9080.286I8
49.3-Methylheptane71.9931.122I8
50.3-Ethylhexane72.1230.058I8
51.trans-1,4-Dimethylcyclohexane72.1230.941N8
52.1-Octene74.6522.155O8
53.1-Ethyl-1-methylcyclopentane74.6520.292N8
54.trans-1,2-Dimethylcyclohexane75.1280.328N8
55.ccc-1,2,3-Trimethylcyclopentane76.0980.159N8
56.n-Octane76.6031.721P8
57.trans-2-Octene77.2150.656O8
58.Isopropylcyclopentane77.4990.261N8
59.cis-2-Octene78.4440.439O8
60.cis-1,2-Dimethylcyclohexane80.0352.055N8
61.1,1,4-Trimethylcyclohexane80.8591.261N9
62.n-Propylcyclopentane82.0270.874N8
63.ctt-1,2,4-Trimethylcyclohexane82.0270.368N9
64.2,5-Dimethylheptane82.2010.733I9
65.3,5-Dimethylheptane82.3980.445I9
66.3,3-Dimethylheptane82.3980.335I9
67.Ethylbenzene83.5011.897A8
68.ccc-1,3,5-Trimethylcyclohexane83.8820.469N9
69.m-Xylene84.720.715A8
70.p-Xylene84.8730.239A8
71.2,3-Dimethylheptane85.0490.22I9
72.3,4-Dimethylheptane85.4110.159I9
73.2-Methyloctane86.2540.245I9
74.ctc-1,2,3-Trimethylcyclohexane86.980.735N9
75.3-Methyloctane87.1211.329I9
76.ctc-1,2,4-Trimethylcyclohexane87.2120.212N9
77.3,3-Diethylpentane87.3890.364I9
78.o-Xylene87.7230.719A8
79.1,1,2-Trimethylcyclohexane87.8290.368N9
80.1-Nonene89.2041.313O9
81.Isobutylcyclopentane89.530.267N9
82.cis-4-Nonene90.0370.829O9
83.trans-3-Nonene90.0370.44O9
84.cis-3-Nonene90.270.575O9
85.n-Nonane90.6411.727P9
86.trans-2-Nonene90.9870.434O9
87.Isopropylbenzene91.9390.48A9
88.2,2-Dimethyloctane92.910.398I10
89.3,3-Dimethyloctane94.9430.153I10
90.n-Propylbenzene95.4070.711A9
91.1-Methyl-3-ethylbenzene96.260.643A9
92.1-Methyl-4-ethylbenzene96.4910.476A9
93.1,3,5-Trimethylbenzene97.1250.097A9
94.2-Methylnonane98.10.3I10
95.1-Methyl-2-ethylbenzene98.20.361A9
96.3-Ethyloctane98.4730.203I10
97.3-Methylnonane98.8071.001I10
98.1,2,4-Trimethylbenzene99.8040.475A9
99.tert-Butylbenzene99.8040.473A10
100.2-Methyl-1-nonene100.0030.648O10
101.Isobutylcyclohexane100.3230.733N10
102.1-Decene100.5540.43O10
103.trans-1-Methyl-2-propylcyclohexane101.460.927N10
104.Isobutylbenzene101.460.713A10
105.n-Decane101.7541.736P10
106.sec-Butylbenzene101.7540.711A10
107.1-Methyl-3-isopropylbenzene102.7860.245A10
108.1-Methyl-4-isopropylbenzene103.1380.949A10
109.1-Methyl-2-isopropylbenzene104.4550.645A10
110.1-Methyl-3-n-propylbenzene105.7380.479A10
111.1-Methyl-4-n-propylbenzene106.1180.477A10
112.n-Butylbenzene106.2120.478A10
113.1,3-Dimethyl-5-ethylbenzene106.3510.094A10
114.1-Methyl-2-n-propylbenzene107.1980.494A10
115.1,4-Dimethyl-2-ethylbenzene108.070.478A10
116.1,2-Dimethyl-4-ethylbenzene108.7470.21A10
117.1,2-Dimethyl-3-ethylbenzene110.4450.495A10
118.1,3-Dimethyl-2-ethylbenzene110.4450.184A10
119.n-Undecane110.7941.723P11
120.1,2,4,5-Tetramethylbenzene111.3940.198A10
121.(2-Methylbutyl)benzene111.560.027A11
122.1-tert-Butyl-2-methylbenzene112.6980.478A11
123.Pentylbenzene114.320.473A11
124.trans-1-Methyl-2-(4-methylpentyl)cyclopentane114.5290.718N12
125.1-t-Butyl-4-ethylbenzene116.0040.478A12
126.n-Dodecane117.6221.727P12
127.1,3,5-Triethylbenzene118.3270.48A12
128.1,2,4-Triethylbenzene119.4840.233A12
129.Hexylbenzene120.6981.186A12
130.n-Tridecane123.1021.723P13
131.5-t-Butyl-1,2,3-trimethylbenzene123.9890.772A13
132.n-Tetradecane127.7191.735P14
133.n-Pentadecane131.7681.734P15
image
GC_PC1330
ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
SampleDHA PiONA standard (cat.# 30730)
Injection
Inj. Vol.:0.2 µL split (split ratio 200:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:0 °C (hold 15 min) to 50 °C at 1 °C/min to 130 °C at 2 °C/min to 270 °C at 4 °C/min
Carrier GasHe, constant pressure
Linear Velocity:24 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
InstrumentAgilent 7890A GC
NotesDetailed hydrocarbon analysis (DHA) is a gas chromatographic technique used to determine the composition of spark ignition fuels. Hydrocarbon streams can be characterized by their individual components or by the group type of hydrocarbon (paraffins = P, isoparaffins = I, olefins = O, naphthenes = N, and aromatics = A). This analysis is a blend of all the group types. To help with identification, individual groups were run under the same conditions and can be found in the following chromatograms:
- GC_PC1322 (isoparaffins)
- GC_PC1323 (aromatic compounds)
- GC_PC1324 (olefins)
- GC_PC1325 (naphthenes)

Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

DHA

,

PiONA

,

PONA

,

gasoline

,

D6729

,

D6730

,

D6733

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us