share:
delim
delim
Permanent Gases on Capillary Rt-MSieve 5A

Permanent Gases on Capillary Rt-MSieve 5A

PeakstR (min)Conc. mol %
1.Hydrogen1.3584
2.Argon1.9763
3.Oxygen2.0282
4.Nitrogen3.5285
5.Methane4.7554
6.Carbon monoxide12.4995
image
GC_PC1331
ColumnRt-Msieve 5A, 30 m, 0.32 mm ID, 30 µm (cat.# 19722)
SamplePermanent gas standard
Diluent:Helium
Injection
Inj. Vol.:50 µL split (split ratio 20:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:40 °C (hold 13 min)
Carrier GasHe, constant flow
Flow Rate:2.5 mL/min
DetectorTCD @ 200 °C
Make-up Gas Flow Rate:10 mL/min
Make-up Gas Type:He
InstrumentAgilent 7890B GC
NotesBaseline resolution between argon and oxygen is obtained using split injection. (R(Ar/O2) = 1.8)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

MolSieve

,

Natural Gas

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us