share:
delim
delim
Hydrocarbon Gases on Rt-Q-BOND
This chromatogram may contain analytes that are synonyms of your search terms.

Hydrocarbon Gases on Rt-Q-BOND

PeaksMol%
1.Methane8.50
2.Ethylene2.50
3.Ethane6.00
4.Propene3.00
5.Propane7.00
6.Propadiene0.85
7.Isobutane6.00
8.Isobutene2.00
PeaksMol%
9.1-Butene2.00
10.1,3-Butadiene3.00
11.n-Butane4.00
12.trans-2-Butene2.00
13.cis-2-Butene2.00
14.Isopentane1.00
15.n-Pentane1.00
16.n-Hexane0.50
image
GC_PC1340
ColumnRt-Q-BOND, 30 m, 0.32 mm ID, 10 µm (cat.# 19744)
SampleRefinery gas standard #1 (cat.# 34441)
Injection
Inj. Vol.:20 µL split (split ratio 20:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:200 °C
Oven
Oven Temp.:40 °C to 250 °C at 10 °C/min
Carrier GasHe, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 260 °C
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Plot Column

,

Refinery Gas

,

PLOT

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us