share:
This chromatogram may contain analytes that are synonyms of your search terms.

Styrene on Stabilwax

Peaks
1.Benzene
2.Toluene
3.Ethylbenzene
4.p-Xylene
5.m-Xylene
6.Cumene
7.o-Xylene
Peaks
8.Propylbenzene
9.p-Ethyltoluene
10.m-Ethyltoluene
11.Styrene
12.Allylbenzene
13.α-Methylstyrene
14.Phenylacetylene
15.trans-β-Methylstyrene
image
GC_PC1352
ColumnStabilwax, 60 m, 0.32 mm ID, 0.25 µm (cat.# 10627)
SampleStyrene
Injection
Inj. Vol.:0.6 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:60 °C (hold 10 min) to 150 °C at 5 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1.7 mL/min
DetectorFID @ 265 °C
InstrumentAgilent 7890B GC
NotesPurity analysis of styrene.
Download PDF