ASTM D7504: Impurities in Monocyclic Aromatic Hydrocarbons on Stabilwax
This chromatogram may contain analytes that are synonyms of your search terms.

ASTM D7504: Impurities in Monocyclic Aromatic Hydrocarbons on Stabilwax

PeakstR (min)
1.Octane4.79
2.Nonane5.60
3.Benzene6.32
4.Decane7.19
5.Toluene8.55
6.1,4-Dioxane9.24
7.Undecane10.48
8.Ethylbenzene11.90
9.p-Xylene12.25
10.m-Xylene12.52
11.Cumene13.88
12.o-Xylene14.37
13.Dodecane14.81
14.Propylbenzene15.35
15.p-Ethyltoluene15.92
PeakstR (min)
16.m-Ethyltoluene16.02
17.tert-Butylbenzene16.49
18.Isobutylbenzene16.62
19.Mesitylene16.77
20.sec-Butylbenzene16.94
21.Styrene17.32
22.Allylbenzene17.60
23.1,2,4-Trimethylbenzene18.22
24.Tridecane18.75
25.1,3-Diethylbenzene18.93
26.1,4-Diethylbenzene19.25
27.n-Butylbenzene19.35
28.α-Methylstyrene20.04
29.Phenylacetylene21.12
30.1,4-Diisopropylbenzene22.25
31.trans-β-Methylstyrene22.46
image
GC_PC1355
ColumnStabilwax, 60 m, 0.32 mm ID, 0.25 µm (cat.# 10627)
SampleD7504 resolution mix
Diluent:Cyclohexane
Injection
Inj. Vol.:0.6 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:60 °C (hold 10 min) to 150 °C at 5 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1.7 mL/min
DetectorFID @ 265 °C
InstrumentAgilent 7890B GC
NotesASTM D7504 covers the determination of trace impurities in monocyclic aromatic hydrocarbons. It includes analysis of nonaromatic and individual aromatic compounds in benzene, toluene, ethylbenzene, p-xylene, o-xylene, styrene, and mixed xylenes.
The chromatogram shows a synthetic blend of the most common analytes identified and quantified by D7504; all analytes are well resolved on a Stabilwax column.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

ASTM D7504

,

Monocyclic Aromatic Hydrocarbons

,

D7504

,

purity test

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us