Determination Of Hydrocarbon Group Type in Gasoline on Rxi-1ms by GC-VUV (ASTM D8071)
Peaks | |
---|---|
1. | Ethanol![]() |
2. | 2-Methylpentane![]() |
3. | Isooctane![]() |
4. | Toluene![]() |
5. | Octane![]() |
Peaks | |
---|---|
6. | Ethylbenzene![]() |
7. | m-Xylene![]() |
8. | p-Xylene![]() |
9. | o-Xylene![]() |
10. | 1,2,4-Trimethylbenzene![]() |
11. | Methylnaphthalenes![]() |

GC_PC1358
Column | Rxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323) |
Sample | Gasoline |
Conc.: | Neat |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 300:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 35 °C (hold 10 min) to 200 °C at 7 °C/min |
Carrier Gas | He, constant flow |
Flow Rate: | 1 mL/min |
Detector | VUV |
Transfer Line Temp.: | 275 °C |
Flow Cell Temp.: | 275 °C |
Acquisition Range: | 125-240 nm |
Acquisition Rate: | 4.5 spectra/sec |
Instrument | HP 6890 GC & VUV Analytics VGA-100 |
Notes | ASTM D8071: Standard Test Method for Determination of Hydrocarbon Group Types and Select Hydrocarbon and Oxygenate Compounds in Automotive Spark-Ignition Engine Fuel Using Gas Chromatography with Vacuum Ultraviolet Absorption Spectroscopy Detection (GC-VUV) |
Acknowledgement | Restek would like to thank Jack Cochran from VUV Analytics for sharing the data. |
RELATED CHROMATOGRAM LIBRARY SEARCHES
ASTM D8071
,D8071
,isoparaffins
,napthenes
,spark igniton engine fuel
,paraffins
,olefins
,aromatics
,gasoline
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.