share:
delim
delim
Jet Fuel on Rxi-1ms by GC-VUV (ASTM D8267)
This chromatogram may contain analytes that are synonyms of your search terms.

Jet Fuel on Rxi-1ms by GC-VUV (ASTM D8267)

image
GC_PC1359
ColumnRxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323)
SampleJet fuel
Conc.:Neat
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:50 °C (hold 0.1 min) to 260 °C at 15 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorVUV
Transfer Line Temp.:275 °C
Flow Cell Temp.:275 °C
Acquisition Range:125-240 nm
Acquisition Rate:7 spectra/sec
InstrumentHP 6890 GC & VUV Analytics VGA-100
NotesASTM D8267, “Standard Test Method for Determination of Saturated Hydrocarbon, Aromatic, and Diaromatic Content of Aviation Turbine Fuels Using Gas Chromatography with Vacuum Ultraviolet Absorption Spectroscopy Detection (GC-VUV),” is used to determine total concentration of saturated hydrocarbon, aromatic, and diaromatic components of aviation turbine engine fuels (jet fuel), not individual compounds.
AcknowledgementRestek would like to thank Jack Cochran from VUV Analytics for sharing the data.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

ASTM D8267

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us