Noble and Permanent Gases on ShinCarbon
This chromatogram may contain analytes that are synonyms of your search terms.

Noble and Permanent Gases on ShinCarbon

Peaks
1.Helium
2.Neon
3.Hydrogen
4.Oxygen
5.Argon
6.Nitrogen
7.Carbon monoxide
8.Methane
9.Krypton
10.Carbon dioxide
11.Xenon
12.Sulfur hexafluoride
image
GC_PC1363
Column100/120 mesh, ShinCarbon ST micropacked column, 2 m, 1.00 mm ID (cat.# 19808)
SampleNoble and permanent gas standard
Conc.:2-10 mol%
Injection
Inj. Vol.:200 µL split (split ratio 10:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:35 °C (hold 3 min) to 80 °C at 10 °C/min to 200 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:6.5 mL/min
DetectorTCD @ 250 °C
InstrumentAgilent 7890B GC
NotesChromatogram A Analysis Conditions:
Inj. Vol.: 50 µL, split (split ratio 10:1)
Conc.: 33 mol% each
Inj. Temp.: 150 °C
Oven Temp.: 35 °C isothermal
Carrier Gas: N2, constant flow
Flow Rate: 5.5 mL/min
Detector: TCD @ 250 °C
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us