Noble and Permanent Gases on ShinCarbon
Peaks | |
---|---|
1. | Helium![]() |
2. | Neon |
3. | Hydrogen![]() |
4. | Oxygen![]() |
5. | Argon![]() |
6. | Nitrogen![]() |
7. | Carbon monoxide![]() |
8. | Methane![]() |
9. | Krypton![]() |
10. | Carbon dioxide![]() |
11. | Xenon![]() |
12. | Sulfur hexafluoride |

GC_PC1363
Column | 100/120 mesh, ShinCarbon ST micropacked column, 2 m, 1.00 mm ID (cat.# 19808) |
Sample | Noble and permanent gas standard |
Conc.: | 2-10 mol% |
Injection | |
Inj. Vol.: | 200 µL split (split ratio 10:1) |
Liner: | Topaz 2.0 mm ID straight inlet liner (cat.# 23313) |
Inj. Temp.: | 150 °C |
Oven | |
Oven Temp.: | 35 °C (hold 3 min) to 80 °C at 10 °C/min to 200 °C at 30 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 6.5 mL/min |
Detector | TCD @ 250 °C |
Instrument | Agilent 7890B GC |
Notes | Chromatogram A Analysis Conditions: Inj. Vol.: 50 µL, split (split ratio 10:1) Conc.: 33 mol% each Inj. Temp.: 150 °C Oven Temp.: 35 °C isothermal Carrier Gas: N2, constant flow Flow Rate: 5.5 mL/min Detector: TCD @ 250 °C |