This chromatogram may contain analytes that are synonyms of your search terms.
Impurities in Biodiesel on Rtx-Biodiesel TG by ASTM D6584
Peaks | tR (min) | |
---|---|---|
1. | Glycerol![]() | 4.80 |
2. | 1,2,4-Butanetriol (IS)![]() | 5.50 |
3. | Monoglycerides | as shown |
Peaks | tR (min) | |
---|---|---|
4. | Tricaprin (IS)![]() | 19.19 |
5. | Diglycerides | 20.16 - 20.9 |
6. | Triglycerides | 22.27 - 25.15 |

GC_PC1367
Column | Rtx-Biodiesel TG with 2 m x 0.53 mm ID retention gap, 15 m, 0.32 mm ID, 0.1 µm (cat.# 10293) |
Sample | Soy B100 biodiesel |
Tricaprin (1,2,3-tricaprinoylglycerol) (cat.# 33025) | |
(s)-(-)-1,2,4-Butanetriol (cat.# 33024) | |
Diluent: | Heptane |
Injection | |
Inj. Vol.: | 1 µL cool on-column |
Temp. Program: | Track oven |
Oven | |
Oven Temp.: | 50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C at 30 °C/min (hold 2 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 2.7 mL/min |
Detector | FID @ 380 °C |
Instrument | Agilent 7890B GC |
Notes | Sample preparation followed ASTM D6584-17: Soy biodiesel B100 was fortified with internal standards (butanetriol and tricaprin) and derivatized using MSTFA. |