share:
delim
delim
Underivatized Ibuprofen on Rt-βDEXsm

Underivatized Ibuprofen on Rt-βDEXsm

image
GC_PH00242
ColumnRt-ßDEXsm, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13104)
Sample
Conc.: 125 ng on-column each enantiomer
Injectionsplit (split ratio 13:1)
Liner:Cup Splitter (cat.# 20709)
Inj. Temp.:200 °C
Oven
Oven Temp.:175 °C to 200 °C at 2 °C/min
Carrier GasHe, constant flow
Linear Velocity:60 cm/sec
DetectorFID @ 230 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

ibuprofen

,

enantiomers

,

NSAID

,

chiral

,

chiral gc separation

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us