share:
delim
delim
Underivatized Barbiturates on Rt-βDEXcst

Underivatized Barbiturates on Rt-βDEXcst

Peaks
1.(+/-)-hexobarbital
2.(+/-)-mephobarbital
image
GC_PH00245
ColumnRt-βDEXcst, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13102)
Sample
Conc.: 25 ng each enantiomer on-column
Injectionsplitless
Oven
Oven Temp.:60 °C (hold 1 min) to 220 °C at 3 °C/min
Carrier GasH2, constant pressure
Linear Velocity:80 cm/sec @ 60 °C
DetectorFID @ 220 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Barbiturates

,

Underivatized

,

Rt-βDEXcst™

,

chiral

,

chiral gc separation

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us