Pure Chromatography
www.restek.com
www.restek.com



This chromatogram may contain analytes that are synonyms of your search terms.
Sympathomimetic Amines (TFAA Derivatives) (Basic Drugs) on Rtx-5
Peaks | |
---|---|
1. | Amphetamine - TFAA![]() |
2. | Phentermine - TFAA![]() |
3. | Phenylpropanolamine - TFAA![]() |
4. | Ephedrine - TFAA![]() |
5. | Methamphetamine - TFAA![]() |
Peaks | |
---|---|
6. | Fenfluramine - TFAA![]() |
7. | Pseudoephedrine - TFAA |
8. | Methylenedioxyamphetamine - TFAA![]() |
9. | Methylenedioxymethamphetamine - TFAA![]() |
10. | Methylphenidate - TFAA![]() |

GC_PH00251
Column | Rtx-5, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10223) |
Sample | Derivatized sympathomimetic amines |
Conc.: | Approximately 2.5 ng/μL |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 1.0 min) |
Inj. Temp.: | 225 °C |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 150 °C at 30 °C/min to 300 °C at 5 °C/min |
Carrier Gas | He, constant pressure |
Linear Velocity: | 20 cm/sec @ 100 °C |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 275 °C |
Ionization Mode: | EI |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.