Sympathomimetic Amines (HFBA Derivatives) (Basic Drugs) on Rtx-5
Peaks | |
---|---|
1. | Amphetamine - HFBA![]() |
2. | Phentermine - HFBA![]() |
3. | Phenylpropanolamine - HFBA![]() |
4. | Ephedrine - HFBA![]() |
5. | Methamphetamine - HFBA![]() |
Peaks | |
---|---|
6. | Fenfluramine - HFBA![]() |
7. | Pseudoephedrine - HFBA |
8. | Methylenedioxyamphetamine - HFBA![]() |
9. | Methylenedioxymethamphetamine - HFBA![]() |
10. | Methylphenidate - HFBA![]() |

GC_PH00252
Column | Rtx-5, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10223) |
Sample | Derivatized sympathomimetic amines |
Conc.: | Approximately 2.5 ng/μL |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 1.0 min) |
Inj. Temp.: | 225 °C |
Oven | |
Oven Temp.: | 40 °C (hold 1 min) to 150 °C at 30 °C/min to 300 °C at 5 °C/min |
Carrier Gas | He, constant pressure |
Linear Velocity: | 20 cm/sec @ 100 °C |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 275 °C |
Ionization Mode: | EI |