share:
delim
delim
Organic Volatile Impurities on Rtx-5 (Rtx-G27)
This chromatogram may contain analytes that are synonyms of your search terms.

Organic Volatile Impurities on Rtx-5 (Rtx-G27)

Peaks
1.Ethylene oxide
2.Methanol
3.Ethanol
4.Diethyl ether
5.1,1-Dichloroethene
6.Acetone
7.Isopropanol
8.Acetonitrile
9.Methylene chloride
10.n-Hexane
11.n-Propanol
12.Methyl ethyl ketone
Peaks
13.Ethyl acetate
14.Tetrahydrofuran
15.Chloroform
16.1,1,1-Trichloroethane
17.Carbon tetrachloride
18.Benzene
19.1,2-Dichloroethane
20.Heptane
21.Trichloroethylene
22.n-Butanol
23.1,4-Dioxane
24.Pyridine
25.Toluene
image
GC_PH00260
ColumnRtx-5 w/5m Integra-Guard Column (Rtx-G27), 30 m, 0.53 mm ID, 5.00 µm (cat.# 10279-126)
SampleHeadspace injection of common solvents for pharmaceutical processing. Prepared to equal about 500 ppm in the bulk pharmaceutical. Samples shaken and heated at 90 °C for 15 minutes, 1 mL headspace injection.
Injection
Inj. Vol.:1,000 µL headspace-syringe split (split ratio 2:1)
Inj. Temp.:220 °C
Oven
Oven Temp.:35 °C (hold 10 min) to 100 °C at 5 °C/min to 240 °C at 25 °C/min (hold 5 min)
Carrier GasHe, constant pressure
Linear Velocity:35 cm/sec @ 35 °C
DetectorFID @ 240 °C
NotesFID sensitivity: 1.05 x 10-11 AFS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Organic

,

Volatile

,

Impurities

,

Rtx®-G27

,

ovi

,

ovis

,

g27

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us