share:
delim
delim
Anesthetics on Rtx-BAC1 & Rtx-BAC2 (headspace analysis)

Anesthetics on Rtx-BAC1 & Rtx-BAC2 (headspace analysis)

Peaks
1.Isoflurane
2.Enflurane
3.Halothane
4.Methoxyflurane
image
GC_PH00501_502
ColumnsRtx-BAC1 30 m, 0.53 mm ID, 3.00 µm (cat.# 18001)
and Rtx-BAC2 30 m, 0.53 mm ID, 2.00 µm (cat.# 18000)
Injection
Inj. Vol.:1,000 µL headspace-syringe split
Inj. Temp.:240 °C
Oven
Oven Temp.:40 °C (hold 5 min) to 240 °C at 5 °C/min
Carrier GasHe, constant flow
Linear Velocity:65 cm/sec
DetectorFID @ 240 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Anesthetics

,

Rtx®-BAC1

,

Rtx®-BAC2

,

analysis

,

headspace

,

bac

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us