γ-Butyrolactone & 1,4-Butanediol Rtx-BAC1 and Rtx-BAC2 (dual column)
Peaks | |
---|---|
1. | 2,3-Butanediol![]() |
2. | 1,2-Butanediol![]() |
3. | 1,3-Butanediol![]() |
4. | 1,4-Butanediol![]() |
5. | γ-Butyrolactone![]() |

GC_PH00808
Columns | Rtx-BAC1 30 m, 0.32 mm ID, 1.80 µm (cat.# 18003) |
and Rtx-BAC2 30 m, 0.32 mm ID, 1.20 µm (cat.# 18002) | |
Sample | |
Diluent: | methanol |
Conc.: | 100 µg/mL each |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 10:1) |
Liner: | Gooseneck Splitless (4mm) w/Wool (cat.# 22405) |
Inj. Temp.: | 240 °C |
Oven | |
Oven Temp.: | 50 °C to 240 °C at 10 °C/min (hold 2 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 3.0 mL/min |
Detector | FID @ 240 °C |