USP Residual Solvents on Rtx-1301 (G43) by Static Headspace
This chromatogram may contain analytes that are synonyms of your search terms.

USP <467> Residual Solvents on Rtx-1301 (G43) by Static Headspace

PeakstR (min)Conc.
(µg/mL)
1.Dichloromethane 5.11012.0
2.Chloroform9.2851.2
3.Benzene 11.1730.04
4.Trichloroethylene 14.6471.6
5.1,4-Dioxane17.4367.6
image
GC_PH00810
ColumnRtx-1301 w/5m Integra-Guard, 30 m, 0.53 mm ID, 3.00 µm (cat.# 16085-126)
SampleUSP <467> Calibration Mixture #5 (cat.# 36007)
Diluent:DMSO
Conc.: To each 22 mL headspace vial 5ml water, ~ 1.0g of sodium sulfate and 100µL of stock standard were added.
Injectionheadspace-loop split (split ratio 2:1)
Headspace-Loop
Inj. Port Temp.:180 °C
Instrument:Teledyne Tekmar HT3
Inj. Time:1.0 min
Transfer Line Temp.:150 °C
Valve Oven Temp.:150 °C
Standby flow rate:10 mL/min
Sample Temp.:80 °C
Platen temp equil. time:2.0 min
Sample Equil. Time:15.0 min
Mixer time:2.0 min
Mixing level:5
Mixer stabilize time:0.5 min
Vial Pressure:15 psi
Pressurize Time:2.0 min
Pressure Equilibration Time:0.5 min
Loop Pressure:5 psi
Loop Fill Time:2.0 min
Loop fill equil. time:0.5 min
Oven
Oven Temp.:40 °C (hold 20 min) to 240 °C at 25 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:5 mL/min
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
NotesFID conditions:
hydrogen flow: 40 mL/min.
air flow: 450 mL/min.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

OVI

,

467

,

USP 467

,

G43

,

Organic volatile impurities

,

OVIs

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us