Pure Chromatography
www.restek.com
www.restek.com

This chromatogram may contain analytes that are synonyms of your search terms.
Basic Drugs (Underivatized) Rxi-5ms
Peaks | |
---|---|
1. | Amphetamine![]() |
2. | Methamphetamine![]() |
3. | Nicotine![]() |
4. | Cotinine![]() |
5. | Caffeine![]() |
6. | Benzphetamine![]() |
Peaks | |
---|---|
7. | Ketamine![]() |
8. | Phencyclidine![]() |
9. | Methadone![]() |
10. | Cocaine![]() |
11. | Codeine![]() |
12. | Scopolamine![]() |
13. | Alprazolam![]() |

GC_PH00814
Column | Rxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) |
Sample | 1000μg/mL each |
Diluent: | methanol |
Conc.: | 20 ng each compound on column |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 50:1) |
Liner: | Siltek Gooseneck Splitless (4mm) (cat.# 20799-214.5) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 220 °C at 15 °C/min to 330 °C at 10 °C/min (hold 5 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 30 cm/sec |
Detector | Agilent 5973 MSD |
Transfer Line Temp.: | 300 °C |
Analyzer Type: | Quadrupole |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Scan Range: | 35-550 amu |