share:
delim
delim
Acidic/Neutral Drugs Rxi-5ms

Acidic/Neutral Drugs Rxi-5ms

Peaks
1.Methprylon
2.Butalbital
3.Amobarbital
4.Meprobamate
5.Glutethimide
6.Phenobarbital
7.Methaqualone
8.Primidone
image
GC_PH00815
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
Sample1000μg/mL each
Diluent:methanol
Conc.:20 ng each compound on column
Injection
Inj. Vol.:1.0 µL split (split ratio 50:1)
Liner:Siltek Gooseneck Splitless (4mm) (cat.# 20799-214.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 220 °C at 15 °C/min to 330 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec
DetectorMS
Mode:Scan
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Tune Type:PFTBA
Ionization Mode:EI
Scan Range:35-550 amu
InstrumentHP6890 GC & 5973 MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

acidic drugs

,

neutral drugs

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us