OVIs by Dynamic Headspace and Gerstel MACH on Rtx-624 (G43) and Rtx-WAX (G16)
This chromatogram may contain analytes that are synonyms of your search terms.

OVIs by Dynamic Headspace and Gerstel MACH on Rtx-624 (G43) and Rtx-WAX (G16)

Peaks
1.Methanol
2.Acetonitrile
3.Dichloromethane
4.trans-1,2-Dichloroethylene
5.Hexane
6.cis-1,2-Dichloroethylene
7.Nitromethane
8.Tetrahydrofuran
9.Chloroform
10.Cyclohexane
11.1,2-Dimethoxyethane
Peaks
12.Trichloroethylene
13.Methylcyclohexane
14.1,4-Dioxane
15.Pyridine
16.Toluene
17.2-Hexanone
18.Chlorobenzene
19.Ethyl benzene
20.m-Xylene
21.p-Xylene
22.o-Xylene
23.Tetralin
* septum components
image
GC_PH00845
ColumnsRtx-624 20 m, 0.18 mm ID, 1.00 µm (cat.# 40924)
and Rtx-WAX (G16) 20m, 0.18mm ID, 0.4µm (custom)
using IP Deactivated Guard Column 5 m, 0.32 mm ID (cat.# 10044)
with Universal “Y” Press-Tight Connector (cat.# 20405)
SampleResidual Solvents Class 2 - Mix A (cat.# 36271)
Residual Solvents Class 2 - Mix B (cat.# 36272)
Diluent:0.5mL DI water/~0.2g sodium sulfate in 20mL headspace vial; components at 0.10 - 6.00µg
Injectionheadspace-trap split (split ratio 20:1)
Headspace-Trap
Instrument:Tekmar HT3
Inj. Temp.:220 °C
Transfer Line Temp.:220 °C
Valve Oven Temp.:220 °C
Standby flow rate:50 mL/min
Trap Standby Temp:40 °C
Trap Sweep Temp:40 °C
Preheat Mixer:On
Preheat Mixer Time:2.0 min
Preheat Mixing Level:5
Preheat Mixer Stabilize Time:0.5 min
Sample Temp.:80 °C
Sweep Flow Rate:75 mL/min
Sweep Flow Time:5.0 min
Dry Purge: 10.0 min, flow 100 mL/min @ 25 °C
Desorb Preheat Temp:245 °C
Desorb: 1.0 min @ 250 °C
Trap Bake Temp:260 °C
Trap Bake Time:6.0 min
Trap Bake Flow:450 mL/min
Oven
Oven Temp.:250 °C
Carrier GasHe
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
InstrumentAgilent 6890 with Gerstel Modular Accelerated Column Heater (MACH)
NotesSample preheat time: 15.00 min.

Flow rate: Constant Flow, Column 1: 0.85mL/min.; Column 2: 0.99mL/min.

Gerstel temps.:

Column 1: 50°C (2 min.) to 80°C at 20°C/min. (1 min.), to 200°C at 40°C/min. (2 min.)
Column 2: 35°C (2 min.) to 60°C at 100°C/min. (1 min.), to 200°C at 40°C/min. (2 min.)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

organic volatile impurity

,

residual solvents

,

G16

,

OVI

,

headspace

,

MACH

,

Gerstel

,

solvents

,

G43

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us