share:
delim
delim
Effect of Liner Choice for Headspace Analysis Residual Solvents Class 2 Mix A on Rtx-1301 (G43)
This chromatogram may contain analytes that are synonyms of your search terms.

Effect of Liner Choice for Headspace Analysis Residual Solvents Class 2 Mix A on Rtx-1301 (G43)

Peaks
1.Acetonitrile
2.Dichloromethane
image
GC_PH00912
ColumnRtx-624, 30 m, 0.32 mm ID, 1.80 µm (cat.# 10970)
SampleResidual Solvents Class 2 - Mix A in 20mL headspace vial (cat.# 36271)
Injectionheadspace-loop split (split ratio 5:1)
Liner:1mm split liner Siltek deactivated and 4mm split IP deactivated liners
Headspace-Loop
Inj. Port Temp.:140 °C
Instrument:Tekmar HT3
Inj. Time:1.00 min
Transfer Line Temp.:105 °C
Valve Oven Temp.:105 °C
Sample Temp.:80 °C
Sample Equil. Time:45 min
Vial Pressure:10 psi
Pressurize Time:0.5 min
Loop Pressure:5 psi
Loop Fill Time:2.00 min
Oven
Oven Temp.:40 °C (hold 20 min) to 240 °C at 10 °C/min (hold 20 min)
Carrier GasHe, constant flow
Flow Rate:2.16 mL/min
Linear Velocity:35.3 cm/sec
DetectorFID @ 240 °C

RELATED CHROMATOGRAM LIBRARY SEARCHES

Class 2

,

headspace

,

OVI

,

G43

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us