This chromatogram may contain analytes that are synonyms of your search terms.

Amphetamines on Rtx-5MS

PeaksTailing Factor
1.Amphetamine1.109
2.Methamphetamine0.992
3.MDA1.106
4.MDMA1.068
5.MDEA1.113
image
GC_PH00972
ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Sampleamphetamine, methamphetamine, MDA, MDMA, and MDEA extracted from methanol and HFAA derivatized
Conc.:100 µg/mL each
Injection
Inj. Vol.:1 µL splitless
Liner:3.5mm custom splitless inlet liner w/IP deactivated wool
Inj. Temp.:220 °C
Oven
Oven Temp.:70 °C (hold 1 min) to 290 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1.25 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:43-450 amu
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us