share:
delim
delim
Residual Solvents Rxi-624Sil MS

Residual Solvents Rxi-624Sil MS

Peaks
1.1,1-Dichloroethene
2.1,1,1-Trichloroethane
Peaks
3.Carbon Tetrachloride
4.Benzene
5.1,2-Dichloroethane
image
GC_PH01115
ColumnRxi-624Sil MS, 30 m, 0.32 mm ID, 1.80 µm (cat.# 13870)
SampleResidual Solvents - Class 1 (cat.# 36279)
Diluent:in 6ml water
Conc.: 2μL in a 20mL headspace vial. (Conc. not consistent with USP <467> methodology.)
Injection
Inj. Vol.:1000 µL headspace-syringe split (split ratio 3:1)
Liner:1mm Split liner (cat.# 20972)
Inj. Temp.:140 °C
Oven
Oven Temp.:40 °C (hold 20 min) to 240 °C at 10 °C/min (hold 20 min)
Carrier GasHe, constant flow
Flow Rate:2.2 mL/min
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
InstrumentAgilent/HP6890 GC
NotesHydrogen: 40mL/min.
Air: 450mL/min.

Note: manual headspace injection
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us