share:
delim
delim
Custom Residual Solvent Mix on Rxi-624Sil MS (using conditions from Pro ezGC)
This chromatogram may contain analytes that are synonyms of your search terms.

Custom Residual Solvent Mix on Rxi-624Sil MS (using conditions from Pro ezGC)

PeakstR (min)Conc.
(µg/mL)
1.Acetonitrile2.056410
2.Dichloromethane2.185600
3.trans-1,2-Dichloroethene2.431940
4.Hexane2.707290
5.Nitromethane3.49750
6.cis-1,2-Dichloroethene3.581940
7.Tetrahydrofuran3.948690
8.Chloroform4.12560
9.1,1,1-Trichloroethane4.3291,500
10.Cyclohexane4.4273,880
11.2-Methoxyethanol5.07950
12.Trichloroethene6.09580
13.Methylcyclohexane6.3731,180
14.1,4-Dioxane6.554380
PeakstR (min)Conc.
(µg/mL)
15.2-Ethoxyethanol6.982160
16.Pyridine7.439200
17.Toluene7.587890
18.Ethylene glycol7.748620
19.2-Hexanone8.26050
20.Dimethylformamide8.597880
21.Chlorobenzene8.882360
22.Ethylbenzene8.985368
23.p-Xylene9.096304
24.m-Xylene9.0961,302
25.o-Xylene9.412196
26.N,N-Dimethylacetamide9.6541,090
27.1-Methyl-2-pyrrolidone11.466530
28.1,2,3,4-Tetrahydronaphthalene12.116100
29.Sulfolane13.266160
image
GC_PH1150
ColumnRxi-624Sil MS, 20 m, 0.18 mm ID, 1.00 µm (cat.# 13865)
SampleResidual solvents mix (cat.# custom)
Injection
Inj. Vol.:0.5 µL split (split ratio 50:1)
Liner:5mm split
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 5 min) to 260 °C at 50 °C/min
Carrier GasHe, constant flow
Flow Rate:1.35 mL/min
Linear Velocity:45 cm/sec
Dead Time:0.74 min @ 40 °C
DetectorFID @ 250 °C
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EZGC

,

Rxi

,

solvents

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us