share:
delim
delim
Column Comparison: Limit of Ethylene Glycol and Diethylene Glycol in Propylene Glycol
This chromatogram may contain analytes that are synonyms of your search terms.

Column Comparison: Limit of Ethylene Glycol and Diethylene Glycol in Propylene Glycol

PeakstR (min)Conc.
(mg/mL)
1.Ethylene glycol3.7570.05
2.Propylene glycol4.4222.0
3.2,2,2-Trichloroethanol (IS)7.4610.1
4.Diethylene glycol10.4160.05
image
GC_PH1151
ColumnRxi-624Sil MS (see notes), 30 m, 0.32 mm ID, 1.80 µm (cat.# 13870)
Sample
Diluent:methanol
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:5mm Single Gooseneck with Wool (cat.# 22973-200.1)
Inj. Temp.:220 °C
Oven
Oven Temp.:100 °C (hold 4 min) to 120 °C at 50 °C/min (hold 10 min) to 220 °C at 50 °C/min (hold 6 min)
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesColumns tested: Rxi-624Sil MS, ZB-624, and DB-624
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

ethylene

,

diethylene

,

glycol

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us