share:
delim
delim
Alkyl Halide Genotoxic Impurities on Rxi-624Sil MS (High Concentration Injection)

Alkyl Halide Genotoxic Impurities on Rxi-624Sil MS (High Concentration Injection)

PeakstR (min)
1.2-Chloropropane2.10
2.Bromoethane2.36
3.1-Chloropropane2.72
4.2-Bromopropane3.393
5.Butyl chloride4.627
6.1-Bromobutane5.973
image
GC_PH1166
ColumnRxi-624Sil MS, 20 m, 0.18 mm ID, 1.00 µm (cat.# 13865)
Sample
Diluent:DMSO
Conc.:1 µg/mL
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:3.5 mm Single Gooseneck Liner with wool placed 3 cm from top (middle) (cat.# 22286)
Inj. Temp.:220 °C
Purge Flow:3 mL/min
Oven
Oven Temp.:40 °C (hold 3 min) to 200 °C at 20 °C/min
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:280 °C
Solvent Delay Time:0.5 min
Ionization Mode:EI
Scan Range:30-300 amu
Scan Rate:5 scans/sec
InstrumentShimadzu 2010 GC & QP2010+ MS
NotesIons displayed: 42, 43, 57, 108 m/z
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Alkyl

,

halide

,

PGI

,

GTI

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us