share:
delim
delim
Residual Solvents Class 2 — Mix A on Rxi-624Sil MS (G43) by USP
This chromatogram may contain analytes that are synonyms of your search terms.

Residual Solvents Class 2 — Mix A on Rxi-624Sil MS (G43) by USP <467>

PeakstR (min)Conc.
(µg/mL)
1.Methanol2.4715.00
2.Acetonitrile4.262.05
3.Methylene chloride4.573.00
4.trans-1,2-Dichloroethene5.074.68
5.cis-1,2-Dichloroethene7.394.68
6.Tetrahydrofuran8.083.60
7.Cyclohexane9.1019.40
8.Methylcyclohexane14.695.90
9.1,4-Dioxane15.691.90
10.Toluene22.414.45
11.Chlorobenzene26.691.80
12.Ethylbenzene27.001.84
13.m-Xylene27.316.51
14.p-Xylene27.321.52
15.o-Xylene28.080.98
16.Cumene28.830.35
Conc. prepared according to USP <467> method.
image
GC_PH1175
ColumnRxi-624Sil MS, 30 m, 0.32 mm ID, 1.80 µm (cat.# 13870)
SampleResidual solvents Class 2 — Mix A (2013 Rev) (cat.# 36012)
Diluent:Water
Injection
Inj. Vol.:1000 µL headspace-loop split (split ratio 5:1)
Liner:Premium 1.0 mm ID straight inlet liner (cat.# 23333.1)
Headspace-Loop
Inj. Port Temp.:140 °C
Instrument:Tekmar HT-3
Inj. Time:1.0 min
Transfer Line Temp.:105 °C
Valve Oven Temp.:105 °C
Needle Temp.:85 °C
Sample Temp.:80 °C
Sample Equil. Time:60 min
Mixer time:5.0 min
Mixing level:5
Mixer stabilize time:0.50 min
Vial Pressure:10 psi
Pressurize Time:5.0 min
Pressure Equilibration Time:0.20 min
Loop Pressure:5.0 psi
Loop Fill Time:2.0 min
Oven
Oven Temp.:40 °C (hold 20 min) to 240 °C at 10 °C/min (hold 20 min)
Carrier GasHe, constant flow
Flow Rate:2.2 mL/min
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

USP467

,

residual solvents

,

,

624

,

solvent

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us