share:
delim
delim
USP Residual Solvents Class 3 MIx A on Rxi-624Sil MS
This chromatogram may contain analytes that are synonyms of your search terms.

USP <467> Residual Solvents Class 3 MIx A on Rxi-624Sil MS

PeakstR (min)Conc.
(µg/mL)
1.n-Pentane2.715000
2.Ethanol2.825000
3.Diethyl ether2.965000
4.Acetone3.315000
5.Ethyl formate3.515000
6.2-Propanol3.525000
7.Methyl acetate3.705000
8.Methyl tert-butyl ether4.125000
9.1-Propanol4.685000
10.2-Butanone5.225000
11.Ethyl acetate5.295000
12.2-Butanol5.485000
13.Isopropyl acetate6.255000
14.Isobutanol6.435000
15.n-Heptane6.775000
16.1-Butanol7.235000
17.Propyl acetate8.015000
18.4-Methyl-2-pentanone9.395000
19.3-Methyl-1-butanol9.655000
20.Isobutyl acetate10.165000
21.1-Pentanol10.945000
22.Butyl acetate11.545000
23.Dimethyl sulfoxide13.245000
24.Anisole13.565000
image
GC_PH1178
ColumnRxi-624Sil MS, 30 m, 0.32 mm ID, 1.80 µm (cat.# 13870)
SampleResidual Solvents Class 3 - Mix A (cat.# 36013)
Conc.:5000 µg/mL split 100:1
Injectionheadspace-loop split (split ratio 100:1)
Liner:Topaz 1 mm Split (cat.# 23333)
Headspace-Loop
Inj. Port Temp.:200 °C
Instrument:Tekmar HT-3
Inj. Time:1.0 min
Transfer Line Temp.:105 °C
Valve Oven Temp.:105 °C
Standby flow rate:50 mL/min
Sample Temp.:80 °C
Platen temp equil. time:80 min
Sample Equil. Time:1.0 min
Mixer time:5.0 min
Mixing level:5
Mixer stabilize time:0.50 min
Vial Pressure:10 psi
Pressurize Time:5.0 min
Pressure Equilibration Time:0.20 min
Loop Pressure:5 psi
Loop Fill Time:2.0 min
Oven
Oven Temp.:35 °C (hold 3 min) to 60 °C at 24 °C/min to 80 °C at 3 °C/min to 280 °C at 25 °C/min
Carrier GasHe, constant flow
Flow Rate:2.7 mL/min
Linear Velocity:41 cm/sec
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us