share:
delim
delim
High Speed GC-MS of Carcinogenic Aryl Amines Resulting From Reductive Cleavage of Azo Dyes on Rxi-35Sil MS (Scan Mode)

High Speed GC-MS of Carcinogenic Aryl Amines Resulting From Reductive Cleavage of Azo Dyes on Rxi-35Sil MS (Scan Mode)

PeakstR (min)Conc.
(µg/mL)
Molecular Ion (m/z)
1.o-Toluidine1.1410107
2.o-Anisidine1.5610123
3.4-Chloroaniline1.7110127
4.p-Cresidine1.9510137
5.2,4,5-Trimethylaniline1.9710135
6.3-Chloro-o-toluidine2.0810141
7.4-Chloro-o-toluidine2.1310141
8.2,4-Diaminotoluene2.7010122
9.2,4,5-Trichloroaniline (IS)3.2530197
10.2-Naphthylamine3.4610143
11.2-Aminobiphenyl3.5010169
12.2-Amino-4-nitrotoluene3.7310152
13.4-Aminobiphenyl4.3110169
14.p-Aminoazobenzene5.6610197
15.4,4'-Oxydianiline5.9010200
16.4,4'-Diaminodiphenylmethane5.9410198
17.Benzidine5.9610184
18.o-Aminoazotoluene6.2310225
19.3,3'-Dimethyl-4,4'-diaminodiphenylmethane6.4810226
20.3,3'-Dimethylbenzidine6.5710212
21.4,4’-Thiodianiline6.9210216
22.3,3'-Dichlorobenzidine7.0510252
23.4,4'-Methylenebis(2-chloroaniline)7.0510266
24.3,3'-Dimethoxybenzidine7.0910244
Elution order of positional isomers was verified with individual standards.
image
GC_QA0096
ColumnRxi-35Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13820)
SampleAccuStandard carcinogenic aryl amine mix (cat.# AE-000-49-R1)
Diluent:Ethyl acetate
Conc.:10 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:100 °C to 320 °C at 27 °C/min (hold 1.75 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.030-3005.5
Transfer Line Temp.:330 °C
Analyzer Type:Quadrupole
Source Type:Inert
Drawout Plate:3 mm ID
Source Temp.:250 °C
Quad Temp.:180 °C
Electron Energy:70 eV
Solvent Delay Time:2.0 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

high speed GC-MS

,

high speed GC

,

fast GC

,

azo dye

,

carcinogenic amines

,

Rxi-35Sil MS

,

textiles

,

regulated azo dyes

,

aryl amines

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us