share:
delim
delim
High Speed GC-MS-SIM of Carcinogenic Aryl Amines Resulting From Reductive Cleavage of Azo Dyes on Rxi-35Sil MS (SIM Mode)

High Speed GC-MS-SIM of Carcinogenic Aryl Amines Resulting From Reductive Cleavage of Azo Dyes on Rxi-35Sil MS (SIM Mode)

PeakstR (min)Quant. Ion (m/z)Qual. Ion #1 (m/z)Qual. Ion #2 (m/z)
1.o-Toluidine1.1410610789
2.o-Anisidine1.5610812380
3.4-Chloroaniline1.71127129100
4.p-Cresidine1.9512213794
5.2,4,5-Trimethylaniline1.97120135134
6.3-Chloro-o-toluidine2.08141106140
7.4-Chloro-o-toluidine2.13141106140
8.2,4-Diaminotoluene2.7012112294
9.2,4,5-Trichloroaniline (IS)3.25195197199
10.2-Naphthylamine3.46143115116
11.2-Aminobiphenyl3.50169168167
12.2-Amino-4-nitrotoluene3.7315210679
13.4-Aminobiphenyl4.31169168167
14.p-Aminoazobenzene5.6692197120
15.4,4'-Oxydianiline5.90200171108
16.4,4'-Diaminodiphenylmethane5.94198197180
17.Benzidine5.9618418592
18.o-Aminoazotoluene6.23106225134
19.3,3'-Dimethyl-4,4'-diaminodiphenylmethane6.48226120211
20.3,3'-Dimethylbenzidine6.57212213106
21.4,4’-Thiodianiline6.92216184183
22.3,3'-Dichlorobenzidine7.05252254256
23.4,4'-Methylenebis(2-chloroaniline)7.05231266195
24.3,3'-Dimethoxybenzidine7.09244201186
image
GC_QA0098
ColumnRxi-35Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13820)
SampleAccuStandard carcinogenic aryl amine mix (cat.# AE-000-49-R1)
Diluent:Ethyl acetate
Conc.:10 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:100 °C to 320 °C at 27 °C/min (hold 1.75 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
11.0089, 106, 107, 123, 80, 108, 127, 129, 100 5
21.82122, 137, 94, 120, 135, 134, 141, 106, 140 5
32.40121, 122, 94, 195, 197, 199 7
43.35143, 115, 116, 169, 168, 167, 152, 106, 79 3
55.0092, 197, 120, 200, 171, 108, 197, 198, 180, 184, 185, 92 3
66.10106, 225, 134, 226, 120, 211, 212, 213, 106 5
76.75216, 184, 183, 252, 254, 256, 231, 266, 195, 244, 201, 186 3
Transfer Line Temp.:330 °C
Analyzer Type:Quadrupole
Source Type:Inert
Drawout Plate:3 mm ID
Source Temp.:250 °C
Quad Temp.:180 °C
Solvent Delay Time:1.0 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

azo dye

,

carcinogenic amines

,

Rxi-35Sil MS

,

textiles

,

regulated azo dyes

,

aryl amines

,

high speed GC-MS

,

high speed GC

,

fast GC

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us