share:
delim
delim
Anti-Ulcerative Cimetidine on Ultra IBD (LC/MS)

Anti-Ulcerative Cimetidine on Ultra IBD (LC/MS)

Peaks
1.Cimetidine
image
LC_0115
ColumnUltra IBD (cat.# 9175552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:ambient
Sample
Diluent:mobile phase
Conc.:10 µg/mL
Inj. Vol.:5 µL
Mobile Phasewater:acetonitrile (20:80, v/v)
Flow:0.2 mL/min
DetectorPE/Sciex API 150 EX
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Ion Spray Voltage:4.8 kV
Orifice Voltage:10 V
Ring Voltage:100 V
Auxiliary Gas:6,000 cc/min.
Source Temp.:350 °C
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us