share:
delim
delim
Phenethyl Glucosinolate on Ultra Aqueous C18

Phenethyl Glucosinolate on Ultra Aqueous C18

Peaks
1.Phenethyl glucosinolate
image
LC_0166
ColumnUltra Aqueous C18 (cat.# 9178565)
Dimensions:150 mm x 4.6 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:ambient
Sample
Diluent:water
Conc.:1,000 µg/mL
Inj. Vol.:10 µL
Mobile Phase
A:50 mM potassium phosphate, pH 2.5
B:acetonitrile
Time (min)%B
0.000
1075
110
160
Flow:1.0 mL/min
DetectorUV/Vis @ 210 nm
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Phenethyl Glucosinolate

,

Ultra Aqueous C18

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us