16-Hr Laboratory Air Sample with radiello 165
This chromatogram may contain analytes that are synonyms of your search terms.

16-Hr Laboratory Air Sample with radiello 165

PeakstR (min)
1.Formaldehyde-DNPH2.80
2.Acetaldehyde-DNPH3.47
3.Acetone-DNPH4.55
4.n-Butyraldehyde-DNPH6.58
5.Hexanal-DNPH10.10
image
LC_AR0508
ColumnRaptor C18 (cat.# 9304A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
SampleLaboratory air sample
Diluent:2 mL acetonitrile extract
Inj. Vol.:5 µL
Mobile Phase
A:Water
B:Methanol:acetonitrile (650:50)*
Time (min)Flow (mL/min)%A%B
0.000.83070
5.000.82575
11.000.81090
11.010.80100
12.000.80100
12.010.83070
14.000.83070
DetectorUV/Vis @ 365, 4.8 nm
Preconcentratorradiello 165
InstrumentWaters ACQUITY UPLC H-Class
NotesThe radiello 165 passive air sampler (RAD165) utilizes a stainless steel net cartridge filled with 2,4-dinitrophenylhydrazine (DNPH) coated Florisil adsorbent. Airborne carbonyls react with the DNPH to form stable hydrazones. The derivatives were subsequently extracted for 30 minutes with 2 mL of acetonitrile. The extract was then analyzed by LC-UV.

*Mobile phase B was prepared by combining 650 mL methanol and 50 mL acetonitrile.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us