share:
delim
delim
Drugs of Abuse on Ultra Biphenyl

Drugs of Abuse on Ultra Biphenyl

PeakstR (min)Positive MRM1Positive MRM2Positive MRM3Negative MRM1Negative MRM2Negative MRM3
1.Ecgonine0.58186.1/168.2186.1/150.9186.1/82.2------
2.Ecgonine methyl ester0.62200.2/82.2200.2/68.2200.2/65.1------
3.Morphine1.35286.1/201.3286.1/229.9286.1/181.1------
4.Amphetamine1.63136.0/91.0136.0/119.1136.0/65.0------
5.Codeine1.76300.2/165.2300.2/152.2300.2/115.2------
6.Methamphetamine1.80150.0/91.2150.0/119.0150.0/65.0------
7.6-MAM1.88328.1/165.1328.1/211.1328.1/152.2------
8.MDMA1.94194.1/163.1194.1/105.1194.1/133.0------
9.Benzoylecgonine2.14290.1/168.1290.1/105.1290.1/77.1------
10.Cocaine2.59304.2/182.1304.2/82.1304.2/105.0------
11.Cocaethylene2.84318.2/196.2318.2/82.2318.2/77.1------
12.PCP3.02244.2/86.1244.2/159.2244.2/91.1------
13.THC-COOH4.27------343.0/299.2343.0/190.8343.0/245.4
14.THC4.63315.3/193.3315.3/93.0315.3/69.1------
The instrument was operated in multi-period mode for this experiment. Period 1 was ESI+, lasting from 0.0 min to 3.5 min, Period 2 was ESI-, lasting from 3.5 min to 4.5 min, Period 3 was ESI+, lasting from 4.5 min to 6.5 min
image
LC_CF0534
ColumnUltra Biphenyl (cat.# 9109552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:30 °C
Sample
Diluent:0.1 Formic acid in water:0.1% formic acid in acetonitrile (98:2)
Conc.:50 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.6982
2.70.66040
4.50.60100
5.00.60100
5.10.6982
DetectorApplied Biosystems/MDS Sciex LC-MS/MS
Model #:API 4000
Ion Source:TurboIonSpray®
Ion Spray Voltage:+/- 4.5 kV
Curtain Gas:40 psi (275.8 kPa)
Gas 1:60 psi (413.7 kPa)
Gas 2:60 psi (413.7 kPa)
Interface Temp.:600 °C
Mode:MRM
InstrumentApplied Biosystems/MDS Sciex LC-MS/MS System
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

NIDA 5

,

morphine

,

Biphenyl

,

drugs of abuse

,

cocaine

,

THC

,

amphetamine

,

opiates

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us