share:
delim
delim
Steroids on Raptor Biphenyl
This chromatogram may contain analytes that are synonyms of your search terms.

Steroids on Raptor Biphenyl

PeakstR (min)Conc.
(µg/mL)
1.Cortisol0.7425
2.Corticosterone1.3125
3.Estradiol1.49100
4.Boldenone1.6625
5.Trenbolone1.8425
6.Testosterone2.0525
7.Methyltestosterone2.2925
8.17-OH-progesterone2.4625
9.Androstenedione3.0225
10.Progesterone4.4825
11.25-Hydroxyvitamin D34.6925
image
LC_CF0559
ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:30 °C
Sample
Diluent:Water:acetonitrile (60:40)
Conc.:25-100 µg/mL
Inj. Vol.:1 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.56040
2.500.55545
5.000.50100
5.010.56040
7.000.56040
DetectorWaters Acquity® PDA @ 254 nm
InstrumentWaters ACQUITY UPLC H-Class
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Raptor Biphenyl

,

raptor

,

steroid

,

biphenyl

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us