share:
delim
delim
Nonsteroidal Anti-Inflammatory Drugs (NSAIDS) on Raptor Biphenyl
This chromatogram may contain analytes that are synonyms of your search terms.

Nonsteroidal Anti-Inflammatory Drugs (NSAIDS) on Raptor Biphenyl

PeakstR (min)Conc.
(µg/mL)
1.Piroxicam1.3925
2.Sulindac1.6025
3.Naproxen1.8525
4.Fenoprofen2.44100
5.Ibuprofen2.54200
6.Diclofenac2.9625
7.Indomethacin3.0825
image
LC_CF0561
ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:40 °C
Sample
Diluent:Water:acetonitrile (65:35)
Conc.:25-200 µg/mL
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.66535
4.000.64060
4.010.66535
5.000.66535
DetectorWater Acquity PDA @ 270 nm
InstrumentWaters ACQUITY UPLC H-Class
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

NSAID

,

raptor biphenyl

,

raptor

,

biphenyl

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us