share:
delim
delim
Nicotine-Related Compounds in Urine on Raptor Biphenyl by LC-MS/MS

Nicotine-Related Compounds in Urine on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion
1.Nornicotine1.25149.1080.05
2.Norcotinine1.81163.0980.05
3.trans-3'-Hydroxycotinine1.88193.1280.05
4.Nicotine1.95163.15132.10
5.Anabasine2.20163.1391.63
6.Cotinine2.88177.1298.07
image
LC_CF0606
ColumnRaptor Biphenyl (cat.# 9309512)
Dimensions:100 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP cartridges 5 mm, 2.1 mm ID, 5 µm (cat.# 930950252)
Temp.:30 °C
Sample
Diluent:Water
Conc.:25 ng/mL (fortified human urine)
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid, 5 mM ammonium formate in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.49010
1.000.49010
2.000.47030
3.000.43070
3.010.49010
5.000.49010
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesThe urine from a non-tobacco user was fortified with analytes to a final concentration of 25 ng/mL and prepared by a liquid-liquid extraction method. A 250 μL urine sample was mixed with 50 μL of 5N NaOH and extracted with 1.5 mL of 50/50 methylene chloride/diethyl ether. A portion of organic phase (1.0 mL) was dried and reconstituted with 200 μL of water for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Nicotine

,

anabasine

,

nicotine metabolite

,

tobacco

,

smoking

,

biphenyl

,

raptor

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us