Hallucinogen Drug Panel on Raptor Biphenyl by LC-MS/MS

Hallucinogen Drug Panel on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.2-Oxo-3-hydroxy-LSD1.21356.3237.0222.0
2.Dextrorphan1.31258.2157.0201.2
3.LSD1.46324.3223.1208.0
4.Dextromethorphan1.57272.2215.1170.9
5.PCP1.60244.286.1159.1
6.25I-NBOMe1.80428.1121.291.1
image
LC_CF0607
ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:50 °C
Sample
Diluent:Water
Conc.:10 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water + 0.1% formic acid
B:Acetonitrile + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.6955
2.000.62575
2.010.6955
4.000.6955
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC