share:
delim
delim
Coenzyme Q10 in Serum on Raptor Biphenyl by LC-MS/MS

Coenzyme Q10 in Serum on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion
1.Ubiquinol-100.98882.82197.18
2.Ubiquinone-10-D91.92872.89206.16
3.Ubiquinone-101.95863.86197.12
image
LC_CF0611
ColumnRaptor Biphenyl (cat.# 9309552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 5 µm (cat.# 930950252)
Temp.:40 °C
Sample
Diluent:See sample preparation
Conc.: Endogenous levels in serum
Inj. Vol.:5 µL
Mobile Phase
A:0.1% formic acid, 5 mM ammonium formate in 50:50 water:acetonitrile
B:0.1% formic acid, 5 mM ammonium formate in methanol
Time (min)Flow (mL/min)%A%B
0.000.6595
2.500.60100
2.510.6595
4.000.6595
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Procedure:
A 150 μL aliquot of serum was mixed with 2.5 μL of ubiquinone-10-D9 (10 μg/mL in methanol) and 850 μL of 1-propanol. Stirred to mix for 2 minutes and then centrifuged at 4,000 rpm for 5 minutes. Transferred a portion of extract to HPLC vial for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

coenzyme

,

coenzyme Q10

,

raptor biphenyl

,

biphenyl

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us