share:
delim
delim
Synthetic Cannabinoids in Urine by LC-MS/MS on Raptor Biphenyl
This chromatogram may contain analytes that are synonyms of your search terms.

Synthetic Cannabinoids in Urine by LC-MS/MS on Raptor Biphenyl

PeakstR (min)Precursor IonQuantifier Product Ion Qualifier Product
1.Pravadoline2.15379.29135.04114.16
2.AM22332.44459.25112.298.15
3.JWH-200-d52.47390.34155.07NA
4.JWH-2002.48385.28155.07114.16
5.WIN 55, 2123.34427.29155.07127.14
6.RCS-44.15322.27135.1277.09
7.XLR-114.21330.25232.17125.1
8.JWH-015-d74.27335.28155.07NA
9.JWH-2504.27336.28121.1291.07
10.JWH-0154.29328.26155.07127.13
11.AM22014.30360.26155.07127.14
12.JWH-2034.39340.23188.18125.09
13.JWH-0734.42328.26155.07127.13
14.UR-1444.44312.32214.17125.1
15.JWH-018-d94.55351.34155.07NA
16.JWH-0184.57342.27155.08127.11
17.JWH-0814.64372.28185.12157.09
18.JWH-1224.69356.29169.12141.11
19.JWH-0194.70356.29155.07127.1
20.JWH-2104.84370.31183.12153.26
image
LC_CF0612
ColumnRaptor Biphenyl (cat.# 9309A5E)
Dimensions:50 mm x 3.0 mm ID
Particle Size:2.7 µm
Guard Column:Raptor Biphenyl EXP guard column cartridge 5.0 mm, 3.0 mm ID, 2.7 µm (cat.# 9309A0253)
Sample
Conc.:5 ng/mL standard was prepared in urine and diluted 3x with 50:50 methanol:water
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.67525
1.000.67525
5.000.6595
5.500.6595
5.510.67525
7.000.67525
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)
Drag and Drop Attachments Here (optional)
...


all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us