share:
delim
delim
Vanillylmandelic Acid, Homovanillic Acid, and 5-Hydroxyindoleacetic Acid in Human Urine (Quantitative Control Level 1) on Raptor Biphenyl

Vanillylmandelic Acid, Homovanillic Acid, and 5-Hydroxyindoleacetic Acid in Human Urine (Quantitative Control Level 1) on Raptor Biphenyl

PeakstR (min)Conc.
(µg/mL)
Precursor IonProduct Ion
1.4-Hydroxy-3-methoxymandelic acid-D3 (VMA-D3)0.900.83200.1139.9
2.Vanillylmandelic acid (VMA)0.922.12–3.90*197.0137.9
3.5-Hydroxyindole-4,6,7-D3-3-acetic-D2 acid (5-HIAA-D5)1.800.83195.1148.0
4.5-Hydroxyindole-3-acetic acid (5-HIAA)1.822.69–4.49*190.0145.9
5.4-Hydroxy-3-methoxyphenyl-D3-acetic-D2 acid (HVA-D5)2.100.83186.1127.0
6.Homovanillic acid (HVA)2.121.33–2.46*181.0121.9
*These concentration ranges were supplied by the vendor (see notes).
image
LC_CF0661
ColumnRaptor Biphenyl (cat.# 9309512)
Dimensions:100 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl guard column cartridge 5 mm, 2.1 mm ID, 5 µm (cat.# 930950252)
Temp.:30 °C
Sample
Diluent:Water
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid, 5 mM ammonium formate in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.58515
3.000.52080
3.010.58515
5.000.58515
DetectorMS/MS
Ion Mode:ESI-
Mode:MRM
InstrumentUHPLC
NotesThe human urine sample is the Bio-Rad Lyphochek quantitative urine control, level 1. A 40 μL aliquot of urine was mixed with 360 μL of water and 10 μL of internal standard solution (33.3 μg/mL in methanol) in a Thomson 0.45 μm PVDF filter vial and injected for analysis after filtration.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

VMA

,

5-HIAA

,

HVA

,

catecholamine metabolite

,

catecholamine

,

serotonin

,

serotonin metabolite

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us