share:
delim
delim
Morphine and Metabolites on Force Biphenyl by UHPLC-MS/MS

Morphine and Metabolites on Force Biphenyl by UHPLC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Morphine-3-β-D-glucuronide0.70462.1286.0152.1
2.Morphine1.26286.1164.9152.0
3.Morphine-6-β-D-glucuronide1.34462.1286.0152.1
4.Morphine-N-oxide1.41302.1284.9161.9
5.6-Acetylmorphine1.86328.1211.1165.0
image
LC_CF0677
ColumnForce Biphenyl (cat.# 9629252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Temp.:35 °C
Sample
Diluent:0.1% Formic acid in water
Conc.:50 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.58515
0.500.58515
2.000.53070
2.010.58515
4.000.58515
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Morphine

,

UHPLC

,

LC-MS/MS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us