share:
delim
delim
Warfarin and Chemotherapeutic Drugs on Raptor FluoroPhenyl by LC-MS/MS

Warfarin and Chemotherapeutic Drugs on Raptor FluoroPhenyl by LC-MS/MS

PeakstR (min)Conc.
(ng/mL)
Precursor IonProduct IonProduct Ion
1.Carboplatin0.83100372.00294.00355.00
2.5-Fluorouracil0.89100130.98113.9957.99
3.Methotrexate2.1820455.15308.09175.14
4.Ifosfamide2.6720261.0391.95153.90
5.Cyclophosphamide2.7520261.03139.97106.01
6.Etoposide3.10100589.34229.01185.03
7.Warfarin3.5810309.11162.99251.01
image
LC_CF0678
ColumnRaptor FluoroPhenyl (cat.# 9319A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor FluoroPhenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9319A0252)
Temp.:50 °C
Sample
Diluent:Water:methanol (95:5)
Conc.:10-100 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:0.1% Formic acid in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.4955
4.000.40100
4.010.4955
6.000.4955
Max Pressure:400 bar
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesThe sample solution was prepared by protein precipitation using Resprep PPT3 96-well plates (cat# 26489) and reservoir (cat# 26493). A 200 μL aliquot of fortified human plasma was mixed with 600 μL of acetonitrile by vortexing for 30 seconds at 2,000 rpm. The sample was filtered using a vacuum manifold. The filtrate was dried and reconstituted with 100 μL of water:methanol (95:5) and injected for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

warfarin

,

Raptor FluoroPhenyl

,

FluoroPhenyl

,

5-fluorouracil

,

methotrexate

,

carboplatin

,

ifosfamide

,

cyclophosphamide

,

etoposide

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us